01
Name the method before drawing the hood
ISO 3679:2022, fifth edition, November 2022, describes three procedures. Rapid equilibrium procedures A and B cover paints, including water-borne paints, varnishes, binders, adhesives, and solvents from −30 °C to 300 °C. Procedure C is a non-equilibrium petroleum test from −20 °C to 300 °C, with precision stated only from 40 °C to 135 °C. The abstract says specifications routinely use A or B. Those sentences decide the cup, the range, and whether a water-borne coating is in the method. They are not a furniture catalogue.
ISO 1523:2002 is a different closed-cup equilibrium method, suitable from about −30 °C to 110 °C depending on the apparatus in its Table 1. It does not apply to water-borne paints and points those to ISO 3679. Clause 7.1 says only to set the apparatus in a draught-free position, preferably in subdued light. It does not mention a fume hood. During heating, the bath and the test portion must not differ by more than 2 °C, and the portion should not rise faster than about 0.5 °C in 1.5 min. Ignition trials are at least 1.5 min apart. The reported flash point is corrected to 101.3 kPa. Halogenated solvent mixtures can give anomalous results. That caution is chemical, not a hood size.
02
The hood sentence is an airflow sentence
ISO 3679:2022 clause 7.2 is the sentence that changes the drawing. Support the apparatus on a level and steady surface in a draught-free position. A draught shield, defined as a shield at the back and on two sides, should be used when protection from draughts is not available. The warning then says that when the material can produce toxic vapours, the apparatus should be located in a fume hood with an individual control of air flow, adjusted so the vapours are withdrawn without causing air currents around the test cup during the test. The word is should. It does not put every coating in a hood, and it does not accept whatever fan happens to be running.
ASTM D3278-21 covers paints, enamels, lacquers, varnishes, solvents, and related products with a flash point between 0 °C and 110 °C and a viscosity lower than 15 000 mm²/s at 25 °C. It uses a 2 mL specimen and a 1 min test. Clause 6.3 defines a draft shield at the back and on two sides where protection from drafts does not exist. The preparation text places the apparatus on a level, stable surface and says not to rely on a laboratory draft hood unless extracted air and vapours can be withdrawn without air currents over the test cup during the ignition. That sentence limits the result. It is not a performance claim for any named hood.
03
A shared fan does not make the cup still
The National Academies' Prudent Practices in the Laboratory, updated 2011, section 9.C.2.7, treats a stopped hood as a different hazard. Keeping chemical hoods running is safer than shutting them to save energy, especially when one hood has its own fan. Under negative pressure, air can be drawn backward through a stopped fan unless an ultralow-leakage backdraft damper is fitted, and cold reverse flow can freeze liquid in the hood. On a constant-air-volume manifold, the book allows 10 to 20 percent of full flow in the off position to limit corrosion. That percentage is a corrosion note. ISO 3679 does not adopt it as an acceptable current over a flash cup.
The two documents therefore refuse the obvious shortcut. The flash method wants no air current around the cup during the test. The laboratory book does not want a single-fan hood switched off in order to buy that still air. Individual airflow control, which ISO 3679 names, is the utility to draw. A room light switch is not that control. The drawing should show a hood that can be turned down for the ignition interval without stopping a fan that other hoods share, and it should say what remains running so the duct does not reverse.
04
Draughts decide whether the face velocity helps
HSE's HSG258, third edition 2017, explains why raising the fan can spoil the cup. Paragraph 104 says draughts reduce hood effectiveness. Causes include nearby turbulence, cooling fans, open doors and windows, and poorly planned make-up air. Paragraph 105 says unsuppressed draughts call for a higher face velocity to limit leakage. A later enclosing-hood note gives a minimum of 0.4 m/s unless a lower velocity is shown to contain the cloud. That figure is HSE design guidance for containment, not a flash-point acceptance limit and not a measured velocity of any furniture on this site.
Paragraph 106 adds the instrument the method implies. An airflow indicator becomes critical when someone must adjust a damper, and HSE says the simplest is usually a manometer. The indicator shows that an adjustment was made. It does not prove the air around the cup was still. HSG258 says draughts can be judged with smoke. The furniture package can lock the hood position, the make-up path, the draught shield, the individual control, and the indicator, with the method number written beside them. It cannot lock a flash point or a statement that the laboratory is compliant.
Before pricing, place the room interface, quantities, and service points for the laboratory fume hood systems with the workflow notes above. Airflow, performance, and compliance stay with the project's responsible reviewers.