Two vessels, one method

Which Assay Acid Can Wait in Polypropylene

A polypropylene shelf is not a second digestion vessel. In the USGS multi-acid method, polypropylene shows up only as a disposable tube for a cooled 1% nitric-acid dilution.

·Industrial testing
Diagram separating a hot Teflon digestion vessel from a polypropylene tube of dilute nitric acid
Conceptual vessel split, not a project photograph. · Original technical illustration — not a project photograph

PROJECT INPUTS

Before a polypropylene item is assigned a duty

  • Keep the 110 °C and 160 °C drying steps in the Teflon vessel named by Open-File Report 99-166
  • Limit the polypropylene tube to the later 0.5 mL plus 4.5 mL dilution in the report's 1% nitric acid
  • Do not treat a 68% nitric-acid cell as a result for 70% acid at 160 °C
  • Do not treat a 40% hydrofluoric-acid cell as clearance for 48% acid in the hot mix
  • Leave 70% perchloric acid, dried at 160 °C, off any shelf justified only by a blank table cell
  • Treat a polypropylene screw cap from a 2.5-minute microwave cycle as a cap, not a cabinet
  • Write chemical, concentration, temperature, and duration for any duty the tables do not contain

01

The method already picks the vessel

In U.S. Geological Survey Open-File Report 99-166 (1999), Briggs and Meier digest 0.200 g for a 42-element ICP-MS method in a 30 mL Teflon vessel, with reference materials, duplicates, and three blanks. The acids are 3 mL hydrochloric acid at 37%, 2 mL nitric acid at 70%, 1 mL perchloric acid at 70%, and 2 mL hydrofluoric acid at 48%. The block is 110 °C to incipient dryness, then 160 °C to complete dryness. Another 1 mL of 70% perchloric acid is dried at 160 °C. After cooling, the vessel gets 1 mL of 70% nitric acid, two drops of 30% hydrogen peroxide, and 19 mL of the report's 1% nitric acid, then 30 minutes at 100 °C while still capped and still Teflon.

Polypropylene enters one step later: 0.5 mL of vessel solution plus 4.5 mL of the same 1% nitric acid, in a capped 13 by 100 mm disposable tube until analysis. The report defines that 1% acid as 10 mL of concentrated nitric acid diluted to 1000 mL. The 100 °C, 30-minute oven step is still the Teflon vessel. No hold time or tube temperature is stated. A shelf for the concentrated bottles, or a tray still on the 160 °C block, is outside this duty.

02

The tables rate other acids and other temperatures

Basell reports ISO 175 tests on 1 mm samples stored 30 days with no load, judged by weight change and an ISO 527-2 tensile test. Plus is resistant, minus is not, and a blank symbol means no test. The note says concentrated nitric acid attacks polypropylene at room temperature. Homopolymer 68% nitric acid is not resistant at 20 °C or 60 °C, with no 100 °C symbol. That cell neither measures nor approves the 70% acid heated to 160 °C in the USGS mixture.

INEOS (February 2012) rates 21 °C, 60 °C, and 100 °C. Ten percent nitric acid is satisfactory at all three. Concentrated nitric acid is some attack at 21 °C and unsatisfactory at 60 °C, with no 100 °C rating, no stated mass percent, and no immersion time. The sheet says a single-reagent line may not hold for a mixture. Basell's 40% hydrofluoric acid is resistant at 20 °C and 60 °C, with no 100 °C symbol; the method uses 48% at 160 °C. Neither guide has 70% perchloric acid at 160 °C. A blank cell is not a pass.

03

A screw cap is not a cabinet

Open-File Report 86-189 puts polypropylene against concentrated acid only as a cap. Samples of 0.100 g, past 100 mesh, went into 250 mL polycarbonate bottles with polypropylene screw caps, Nalge 3122-0250, threads wrapped twice in Teflon tape. Each bottle took 2.0 mL concentrated nitric acid and 5.0 mL of a 7:3 hydrochloric-hydrofluoric mix. Twelve bottles ran at 650 W for 2.5 minutes, then a 2-minute air purge, 10 minutes in ice water, 93.0 mL of 1.5% boric acid (30.0 g orthoboric acid in 2 L), and another 10 minutes at 650 W. The oven sat in a fume hood.

Acid fumes may leak around those caps, and the report calls that a health hazard unless the oven vents to a hood. The 135 °C figure is the polycarbonate melting point, a limit on bottle contents if time rises or cavity volume falls, not a polypropylene shelf rating. Bottle life was typically four runs, with yellowing and frosting called normal. Recoveries exceeded 95% for most elements in most of 51 materials, including rocks, sediments, sulfide ores, and soils; aluminum, chromium, lithium, lead, silicon, and zirconium were low when refractory minerals were present. A leaking cap from those timed cycles is not a cabinet test.

04

What the tube may hold

The polypropylene duty in Open-File Report 99-166 is the capped 13 by 100 mm tube: 0.5 mL of vessel solution plus 4.5 mL of the report's 1% nitric acid. That is not 70% nitric acid, 48% hydrofluoric acid, dried 70% perchloric acid, or the 160 °C block. INEOS rates 10% nitric acid satisfactory at 21 °C, 60 °C, and 100 °C, but 1% is below that row and the method states no tube temperature. Two drops of 30% hydrogen peroxide stay in the cooled Teflon vessel. Basell calls 30% peroxide resistant at 20 °C and limited, with noticeable swelling, at 60 °C. INEOS marks some attack at 60 °C.

Concentrated acids, the mixed digestion, perchloric residues dried at 160 °C, and anything still at 110 °C or 160 °C stay in the Teflon vessel. Basell rates 36% hydrochloric acid resistant at 20 °C and 60 °C, with no 100 °C symbol. The method uses 37% and then heats it with the other three acids, so that row does not clear the mix. A polypropylene cabinet around the bottles needs its own chemical, concentration, temperature, and duration. This note supplies none of that test and reports no result for a furnished cabinet. The usable rule is the method's vessel split, read only as far as the printed symbols.

Where the storage duty is considering a polymer cabinet or bench, review PP laboratory furniture on its own. A material name does not show that it fits the chemicals, cleaning method, or fire requirements.

SOURCE REVIEW

Reviewed sources

Sources support the planning context. The article is original YOJIA content and does not reproduce source publications.

  1. PP Chemical Resistance Guide, February 2012INEOS Olefins & Polymers USA · 2026-11-16

RFQ / Project inquiry

Turn the application article into a room-level brief

Share the room plan, equipment list, user workflow, destination, and known requirements so the furniture scope and interfaces can be reviewed.